  SPELLMAN高压电源广泛应用于东芝CT机XVISION/EX、AUKLET系列、ASTEION系列、以及多排系列。故障代码及诊断方法大致相同,部分故障代码排查参考如下

  Error contents (note 1)

  



  Cause of problem

  


  Object part

  





  24kW

  


  48kW/60kW

  


  E13

  


  Arc-

  


  Cathode side Arc of the X-ray tube or the high-voltage section

  


  TUBE, INV/HV

  


  TUBE, INV/HVC

  


  E25

  


  Arc+

  


  Anode side Arc of the X-ray tube or the high-voltage section

  



  TUBE, INV/HVA

  


  E12

  


  Over kV-

  


  Cathode side overvoltage. Abnormal main circuit control.

  


  INV/HV, SCB

  


  INV/HVC, SCB

  


  E27

  


  Over kV+

  


  Anode side overvoltage. Abnormal main circuit control.

  



  INV/HVA, SCB

  


  E62

  


  Over mA-

  


  Cathode side overcurrent of the X-ray tube or the high-voltage section.

  


  TUBE, INV/HV, SCB

  


  TUBE, INV/HVC,SCB

  


  E63

  


  Over mA+

  


  Anode side overcurrent of the X-ray tube or the high-voltage section.

  



  TUBE, INV/HVA,SCB

  


  E66

  


  MAC error

  


  Filament current abnormal. Trouble of the filament heatig circuit.

  


  INV/HV, SCB

  


  INV/HVC, SCB

  


  E0

  


  Regulation error (HV abnormal)

  


  There is a difference in binding output and real output

  


  INV/HV, AC/C

  


  INV/HVC, INV/HVA, AC/C

  