芝CT机spellman电源维修CT42 CT70北京
SPELLMAN高压电源广泛应用于东芝CT机XVISION/EX、AUKLET系列、ASTEION系列、以及多排系列。故障代码及诊断方法大致相同,部分故障代码排查参考如下
Error contents (note 1)
| Cause of problem
| Object part
| ||
24kW
| 48kW/60kW
| |||
E13
| Arc-
| Cathode side Arc of the X-ray tube or the high-voltage section
| TUBE, INV/HV
| TUBE, INV/HVC
|
E25
| Arc+
| Anode side Arc of the X-ray tube or the high-voltage section
| TUBE, INV/HVA
| |
E12
| Over kV-
| Cathode side overvoltage. Abnormal main circuit control.
| INV/HV, SCB
| INV/HVC, SCB
|
E27
| Over kV+
| Anode side overvoltage. Abnormal main circuit control.
| INV/HVA, SCB
| |
E62
| Over mA-
| Cathode side overcurrent of the X-ray tube or the high-voltage section.
| TUBE, INV/HV, SCB
| TUBE, INV/HVC,SCB
|
E63
| Over mA+
| Anode side overcurrent of the X-ray tube or the high-voltage section.
| TUBE, INV/HVA,SCB
| |
E66
| MAC error
| Filament current abnormal. Trouble of the filament heatig circuit.
| INV/HV, SCB
| INV/HVC, SCB
|
E0
| Regulation error (HV abnormal)
| There is a difference in binding output and real output
| INV/HV, AC/C
| INV/HVC, INV/HVA, AC/C
|
Spellman高压电源XRM/PCM50/70维修射频电源
Spellman高压电源XRM50P50X3768维修射频电源
spellman高压电源维修高压发生器控制单元维修